header Sub PASAN

CTLab 606

The accuracy of the power measurement is set by the quality and reliability of the tester. The IEC Standard IEC 60904-9 classifies sun simulators / cell testers in three classes A (the best), B and C depending on three parameters:

 

IEC 60904-9Class ACTLab 606
Non-uniformity of irradiance≤ 2 %≤ 1.0 %
Pulse instability (long term)≤ 2 %≤ 1.0 %
Spectral irradiance distribution≤ ± 25 %≤ ± 25 %

This means that the CTLab 606 could be rated class AA as regards light uniformity and pulse stability and class A as regards spectral match. This outstanding quality is the result of Pasan’s more than 25 years of experience and technical expertise in the field of PV measurement.


The CTLab 606 meets the highest requirements of R&D-labs The Xenon-flash based equipment is fitted for temperature dependent measurements of all relevant IV parameters. Cells up to 22 x 22 cm (8”) with two or three busbars can be measured. Also, part of the equipment is Pasan’s unique long-lasting and robust contact system, which casts only a minor shadow thus reducing any interference with the light source.

Main features

  • Better than IEC class AAA
  • Cell size up to 8.8 inches (22 x 22 cm)
  • Pulse duration 4 ms
  • Temperature range from 15 to 60 °C
  • Fully automated temperature coefficient determination

System includes

  • Electronic load with flash generator
  • Flash box assembly including chuck and contact frame
  • Temperature-adjustable chuck
  • Monitor cell
  • PC with operating software