전체 제품
대표 플랫폼
전체 라인업
살펴보기
문의하기

Actual issues on power measurement of photovoltaic modules

Paul Grunow, Alexander Preiss, Michael Schoppa, Stefan Krauter

Institution
Photovoltaik Institut Berlin (PI Berlin), Germany
Equipment named
Pasan SSIIIb (class A, 3.0 m x 3.0 m)
Application
Reference simulator at PI Berlin in a study of uniformity, spectral mismatch, tandem current mismatch and sweep-time (transient) effects on Pmax of crystalline and thin-film modules (a-Si, micromorph tandem, CIGS).
Use
Accredited testing
Published uncertainty
Pmax ±2.2% (k = 2) for single-junction modules and 2.9% (k = 2) for tandem modules; measured uniformity ±0.3% on 3 m x 3 m
Technologies
Crystalline silicon, Thin film, Multi-junction, CIGS
Venue
Photovoltaics International, 8th print edition, 2010

Significance

Independent test lab characterising its own Pasan SSIIIb: +/-0.3% uniformity on 3x3 m against the +/-2% class A limit.

From the source

“The measured uniformity of ±0.3% on a Pasan SSIIIb achieved that requirement very easily. [...] The maximum sweep time of the Pasan SSIIIb is 8ms, which proved to be sufficient for standard a-Si, mc-Si, and sc-Si modules (max. deviation of 0.5%).”

Read the source

This entry summarises a third-party publication. It does not imply that the institution or the authors endorse Pasan SA or its products.