Power Measurement Analysis of Micromorph Tandem Modules

Johannes Meier et al.

Institution
TEL Solar metrology laboratory ('the company's metrology laboratory'); author affiliations TEL Solar-Lab SA (Neuchâtel), TEL Solar AG (Trübbach), Photovoltaic Consultant (Freiburg), Switzerland (author affiliations)
Equipment named
modified solar simulator Pasan SunSim IIIb; Pasan electrical load with 16 bit data acquisition
Application
STC power measurement and calibration of Micromorph reference modules, a 'fading' multi-flash procedure for transient effects, and spectrometric assessment of tandem current matching
Use
Secondary calibration
Published uncertainty
Expanded uncertainty (k=2) of 2.9% (2.89%) on Pmax at STC for Micromorph reference module calibration
Technologies
Thin film, Multi-junction
Venue
Conference paper, 28th European Photovoltaic Solar Energy Conference and Exhibition (EU PVSEC 2013), pp. 3513-3518, 2013

Significance

Completes the uncertainty analysis of the tandem set-up and introduces an advanced multi-flash method, qualifying reference modules at 2.9% (k=2).

From the source

“The modified solar simulator Pasan SunSim IIIb, its spectral adjustment and the method of using traceable reference component cells were published in ref. 1.”

Read the source

This entry summarises a third-party publication. It does not imply that the institution or the authors endorse Pasan SA or its products.