Implementation of the IEC TS 60904-1-2 Measurement Methods for Bifacial Silicon PV Devices

Juan Lopez-Garcia et al.

Institution
European Commission JRC, ESTI, Ispra, Italy
Equipment named
class AAA PASAN IIIB PSS (pulse length of 10 ms)
Application
Single-side illumination I-V measurements of bifacial devices (multiflash method) for IEC TS 60904-1-2 implementation.
Use
Research
Published uncertainty
Reference cell used with the PSS: certified UC (k=2) of +/-0.48% (reference cell, not the simulator); nonuniformity 0.45% on 3x3 m2
Technologies
Bifacial
Venue
IEEE Journal of Photovoltaics, 2022

Significance

Shows the ESTI Pasan IIIB used for pre-normative bifacial measurement methods; reports 0.45% nonuniformity over 3x3 m2 and <0.1% flash-to-flash irradiance change.

From the source

“I-V curve measurements were carried out at ESTI facilities using a class AAA PASAN IIIB PSS (pulse length of 10 ms) with the so-called multiflash (MF) method”