All products
Flagship platforms
Full range
Explore
Get in touch

Measuring CdTe thin-film modules: current-injection preconditioning, step by step

CdTe modules are metastable: the power they show after storage in the dark is not the power they deliver in the field. Why that is, what the standards and the research say about preconditioning them, a practical current-injection sequence for a flasher, and the mistakes that most often make two labs disagree on the same module.

Measuring CdTe thin-film modules: current-injection preconditioning, step by step

Cadmium telluride (CdTe) is the most widely installed thin-film technology, and CdTe modules reach test laboratories from factories, from power plants and from round robins. They are measured on the same flashers as crystalline silicon, but they do not behave like crystalline silicon, and a flash taken the way it would be for a TOPCon module can be several percent off.

Why CdTe needs preconditioning

CdTe devices are metastable: their I-V curve depends on their recent history of light, electrical bias and temperature. After storage in the dark the module typically loses fill factor and open-circuit voltage. Light or a forward current brings it back to its operating state, and once the exposure stops it slowly relaxes again. A module flashed straight out of a dark store is therefore measured in a state it never has in the field, and two laboratories that handle it differently will not agree.

Two ways to condition: light or current

  • Light soaking is the standard route. IEC 61215 stabilises thin-film modules (MQT 19, with the CdTe-specific requirements in IEC 61215-1-4) under light, in intervals of at least 10 kWh/m², until the power changes by less than 2 % from one interval to the next. It is the reference for type approval, but it takes hours per module and a light soaker.
  • Forward current in the dark drives the module as a diode from an external supply. Work at NREL between 2009 and 2011 showed that dark forward bias, like light soaking, brings CdTe modules to a stable state, provided the current is at least of the order of the short-circuit current. Module makers describe the same mechanism in their patents: First Solar’s US 9,202,964 gives a biasing window of 0.3 to 5 times Isc for 1 to 60 minutes.

Current injection takes minutes instead of hours, needs only a DC supply and a switch, and is done on the flasher itself, right before the flash. That is why it is used before the power measurement of CdTe modules in production lines and in test laboratories. For a type-approval result, follow the stabilisation of IEC 61215, or report the method you used.

A practical sequence on a flasher

  • Bring the module to the test temperature, 25 °C, before injecting. The module heats up a little during the injection, so check its temperature again just before the flash.
  • Connect the supply through a switch box that disconnects it from the module before the flash, so that the electronic load sees only the module during the I-V sweep.
  • Ramp the current up over about 0.1 to 1 s instead of switching it on in one step. It protects both the supply and the module.
  • Hold a set current for a set time. The higher the current, the shorter the hold. As starting points: about 1 × Imp for 4 to 6 minutes, 2 × Imp for 1 to 3 minutes, 3 × Imp for 40 to 120 seconds, or 4 × Imp for 20 to 60 seconds. When the module maker gives its own procedure, use it.
  • Release the current and flash within a fraction of a second, a few hundred milliseconds at most, and keep that delay identical for every module: the module starts to relax the moment the current stops.
  • Repeat the whole sequence on the same module until its maximum power stops changing within your repeatability. That repeat test, not the recipe, is what proves your settings are sufficient.

Size the supply for the module. A CdTe module is a high-voltage, low-current device, and pushing the injection current through it in the forward direction takes a voltage somewhat above its open-circuit voltage. Use a current-limited supply rated above Voc with margin, and treat the set-up as a high-voltage installation.

The other reasons two labs disagree

  • Reference device and spectral mismatch. CdTe responds only up to about 850 to 900 nm, a much narrower band than silicon. Calibrate with a CdTe reference module, or apply the spectral mismatch correction of IEC 60904-7 with the measured spectral responses of module and reference. A silicon reference without correction biases the power.
  • A large calibration correction. If calibrating on the reference module needs a correction of several percent, something is wrong in the set-up or the reference: find out what before measuring modules.
  • Timing and handling. The delay between conditioning and flash, the module temperature, and whether the module is moved or stored in between must be the same for every module and written into the procedure.

How Pasan systems do it

Current-injection preconditioning is available as an option on Pasan HighLIGHT xenon simulators such as the HighLIGHT 100ms Lab and HighLIGHT 100ms Prod, and on the Nexun LED simulators of our sister company Avalon ST. A precon switch box connects your DC supply to the module, Pasan Studio holds the injection for the time you set, releases it and fires the flash straight after, and the preconditioning time is stored with the measurement. For light-soak stabilisation, the LED Light Soaker holds a full-size module under regulated continuous light.

Send us a CdTe module or ask for a demonstration, and read what light soaking does and when you need it.

References

  • J. A. del Cueto et al., “Striving for a standard protocol for preconditioning or stabilization of polycrystalline thin film photovoltaic modules”, Proc. SPIE 7412, 2009. DOI 10.1117/12.825380
  • C. Deline, J. A. del Cueto, D. S. Albin, S. R. Rummel, “Metastable electrical characteristics of polycrystalline thin-film photovoltaic modules upon exposure and stabilization”, Proc. SPIE 8112, 2011. DOI 10.1117/12.893813
  • T. J. Silverman, M. G. Deceglie, B. Marion, S. R. Kurtz, “Performance stabilization of CdTe PV modules using bias and light”, IEEE Journal of Photovoltaics, 2015. DOI 10.1109/JPHOTOV.2014.2370252
  • First Solar, US 9,202,964 B2, “System and method for photovoltaic device temperature control while conditioning a photovoltaic device”, 2015, and US 8,940,556 B2, “Electrical bias methods and apparatus for photovoltaic device manufacture”, 2015.
  • IEC 61215-1-4 and IEC 61215-2 (MQT 19, stabilisation); IEC 60904-7 (spectral mismatch correction).
  • IEA-PVPS T13-02:2014, “Characterisation of performance of thin-film photovoltaic technologies”.
Pasan
The solar reference since 1978
Related products
HighLIGHT 100ms LabHighLIGHT 100ms ProdLED Light Soaker

Rather we contact you?

Leave your email and we will get back to you. No form to fill in.

We only use your address to answer you. Privacy policy

Talk to us

Working on something similar? Let’s talk.

If you’re tackling a measurement challenge that resembles what you’ve just read about, send us a note. We answer email.