The measurement problem
A power value without its uncertainty and traceability cannot be compared with a label, another laboratory or a degradation threshold. Two shortcuts mislead buyers: quoting repeatability (how stable a system is) as if it were accuracy (how close it is to the SI value), and quoting a class letter as if it were an uncertainty.
The physics behind it
| Contributor | Published magnitude | Source |
|---|---|---|
| Reference device calibration (primary level, k = 2) | 0.23 % (multiply calibrated set) to 0.4 % (laser DSR) | [1][2] |
| Spectral mismatch factor uncertainty | 0.06 % to 1.26 % (k = 2), by correlation assumptions | [5] |
| Uncorrected mismatch, c-Si vs c-Si on AAA simulators | About 2 % | [3] |
| Capacitive error, step-like methods | (0.47 ± 0.19) % (k = 2) | [3] |
| Non-uniformity | Often below 1 % over a wide range of values, case by case | [6] |
| Preconditioning (thin film) | Dominant cause of inconsistencies between reference laboratories | [7] |
| Overall module Pmax, accredited | ±1.1 % monofacial, ±1.4 % bifacial (coverage not stated on the page) | [4] |
| Interlaboratory agreement, reference laboratories (2017) | c-Si within ±0.5 % (all within ±1 %); thin film ±1 to 2 % | [7] |
What the measurement must achieve
- A traceable reference device with a calibration certificate stating its uncertainty.
- Mismatch correction with measured spectral data, or a matched reference.
- Uniformity, stability and spectrum characterised at the operating conditions, not only at acceptance.
- A method shown equivalent to steady state for the device.
- Controlled preconditioning and temperature.
- A GUM uncertainty budget per technology, and participation in intercomparisons.
Applicable standards
| Standard | What it governs here |
|---|---|
| JCGM 100:2008 (GUM)Evaluation of measurement data: Guide to the expression of uncertainty in measurement | Method for evaluating and expressing uncertainty. |
| ISO/IEC 17025:2017General requirements for the competence of testing and calibration laboratories | Traceability (6.5), uncertainty evaluation (7.6), validity of results (7.7). |
| IEC 60904-1:2020Measurement of photovoltaic current-voltage characteristics | I-V measurement and data analysis. |
| IEC 60904-7:2019Computation of the spectral mismatch correction | Spectral mismatch. |
| IEC 60904-9:2020 (edition 3)Classification of solar simulator characteristics | Classification, which by its own statement does not predict power uncertainty. |
Measurement technologies: advantages and limitations
The dominant terms depend on method and device more than on lamp type. Pulsed systems minimise device heating; steady-state systems need temperature control; LED systems allow spectral tuning that can reduce mismatch but add the uncertainty of each channel's calibration [8]. No study was found that ranks lamp technologies by total module power uncertainty.
Where Pasan equipment fits
- HighLIGHT A++ ReferencePasan's reference simulator: spectral match within ±6.25 %, non-uniformity below 0.5 %, long-term instability below 0.5 % (Pasan's A++ convention: half of the IEC A+ limits), traceable to METAS.
- WPVS transimpedance amplifierReads non-shunted WPVS reference cells with ±0.1 % accuracy at 25 °C.
- Spectral Response KitSpectral data for the mismatch correction.
Technical questions
- Is a repeatability of 0.05 % an accuracy of 0.05 %?
- No. Repeatability is agreement under identical conditions; absolute uncertainty includes the reference calibration, 0.23 to 0.4 % at the primary level alone [1][2].
- What is the best published module power uncertainty?
- Fraunhofer ISE CalLab states ±1.1 % for monofacial modules (accredited) [4].
References
- , “Reduction of uncertainties for photovoltaic reference cells”, Metrologia 52, 646-653, 2015. doi:10.1088/0026-1394/52/5/646
- , “Design, realization and uncertainty analysis of a laser-based primary calibration facility for solar cells at PTB”, Measurement 51, 457-463, 2014. doi:10.1016/j.measurement.2013.12.001
- , “Electrical characterization intercomparison of high-efficiency c-Si modules within Asian and European laboratories”, Progress in Photovoltaics 27, 603-622, 2019. doi:10.1002/pip.3134
- , “CalLab PV Modules (accredited calibration laboratory), web page accessed November 2026”, Fraunhofer ISE, 2026. Source
- , “Spectral mismatch uncertainty estimation in solar cell calibration using Monte Carlo simulation”, IEEE Journal of Photovoltaics 13, 899-904, 2023. doi:10.1109/JPHOTOV.2023.3311890
- , “IEC 60904-9:2020, Photovoltaic devices, Part 9: Classification of solar simulator characteristics (edition 3.0)”, International Electrotechnical Commission, 2020. Source
- , “Improvements in world-wide intercomparison of PV module calibration”, Solar Energy 155, 1451-1461, 2017. doi:10.1016/j.solener.2017.07.081
- , “Traceable efficiency determination of large-area perovskite on silicon tandem modules: insights from a calibration laboratory's perspective”, Progress in Photovoltaics 33, 815-827, 2025. doi:10.1002/pip.3928
- , “Analysis and mitigation of measurement uncertainties in the traceability chain for the calibration of photovoltaic devices”, Measurement Science and Technology 20, 075101, 2009. doi:10.1088/0957-0233/20/7/075101
How this page is written: claims about physics and standards cite published sources, claims about Pasan equipment repeat only what the product pages state, and no other manufacturer is named. Editorial method