High-accuracy PV power measurement: what actually limits it

The accuracy of a module power measurement is limited by a few terms whose sizes are published: the calibration of the reference device (0.23 to 0.4 % at the primary level, k = 2 [1][2]), spectral mismatch, non-uniformity, capacitive error ((0.47 ± 0.19) % residual with step-like methods [3]), preconditioning for thin film, and temperature; the best accredited module calibrations today state about ±1.1 % [4], and neither a repeatability figure nor a simulator class measures accuracy.

The measurement problem

A power value without its uncertainty and traceability cannot be compared with a label, another laboratory or a degradation threshold. Two shortcuts mislead buyers: quoting repeatability (how stable a system is) as if it were accuracy (how close it is to the SI value), and quoting a class letter as if it were an uncertainty.

The physics behind it

Typical contributors with published examples
ContributorPublished magnitudeSource
Reference device calibration (primary level, k = 2)0.23 % (multiply calibrated set) to 0.4 % (laser DSR)[1][2]
Spectral mismatch factor uncertainty0.06 % to 1.26 % (k = 2), by correlation assumptions[5]
Uncorrected mismatch, c-Si vs c-Si on AAA simulatorsAbout 2 %[3]
Capacitive error, step-like methods(0.47 ± 0.19) % (k = 2)[3]
Non-uniformityOften below 1 % over a wide range of values, case by case[6]
Preconditioning (thin film)Dominant cause of inconsistencies between reference laboratories[7]
Overall module Pmax, accredited±1.1 % monofacial, ±1.4 % bifacial (coverage not stated on the page)[4]
Interlaboratory agreement, reference laboratories (2017)c-Si within ±0.5 % (all within ±1 %); thin film ±1 to 2 %[7]

What the measurement must achieve

  • A traceable reference device with a calibration certificate stating its uncertainty.
  • Mismatch correction with measured spectral data, or a matched reference.
  • Uniformity, stability and spectrum characterised at the operating conditions, not only at acceptance.
  • A method shown equivalent to steady state for the device.
  • Controlled preconditioning and temperature.
  • A GUM uncertainty budget per technology, and participation in intercomparisons.

Applicable standards

Applicable standards
StandardWhat it governs here
JCGM 100:2008 (GUM)Evaluation of measurement data: Guide to the expression of uncertainty in measurementMethod for evaluating and expressing uncertainty.
ISO/IEC 17025:2017General requirements for the competence of testing and calibration laboratoriesTraceability (6.5), uncertainty evaluation (7.6), validity of results (7.7).
IEC 60904-1:2020Measurement of photovoltaic current-voltage characteristicsI-V measurement and data analysis.
IEC 60904-7:2019Computation of the spectral mismatch correctionSpectral mismatch.
IEC 60904-9:2020 (edition 3)Classification of solar simulator characteristicsClassification, which by its own statement does not predict power uncertainty.

Measurement technologies: advantages and limitations

The dominant terms depend on method and device more than on lamp type. Pulsed systems minimise device heating; steady-state systems need temperature control; LED systems allow spectral tuning that can reduce mismatch but add the uncertainty of each channel's calibration [8]. No study was found that ranks lamp technologies by total module power uncertainty.

Where Pasan equipment fits

  • HighLIGHT A++ ReferencePasan's reference simulator: spectral match within ±6.25 %, non-uniformity below 0.5 %, long-term instability below 0.5 % (Pasan's A++ convention: half of the IEC A+ limits), traceable to METAS.
  • WPVS transimpedance amplifierReads non-shunted WPVS reference cells with ±0.1 % accuracy at 25 °C.
  • Spectral Response KitSpectral data for the mismatch correction.

Technical questions

Is a repeatability of 0.05 % an accuracy of 0.05 %?
No. Repeatability is agreement under identical conditions; absolute uncertainty includes the reference calibration, 0.23 to 0.4 % at the primary level alone [1][2].
What is the best published module power uncertainty?
Fraunhofer ISE CalLab states ±1.1 % for monofacial modules (accredited) [4].
Which term usually dominates?
For c-Si, reference calibration and spectral mismatch [9]; for thin film, preconditioning [7].

References

  1. H. Müllejans et al., “Reduction of uncertainties for photovoltaic reference cells”, Metrologia 52, 646-653, 2015. doi:10.1088/0026-1394/52/5/646
  2. S. Winter et al., “Design, realization and uncertainty analysis of a laser-based primary calibration facility for solar cells at PTB”, Measurement 51, 457-463, 2014. doi:10.1016/j.measurement.2013.12.001
  3. C. Monokroussos et al., “Electrical characterization intercomparison of high-efficiency c-Si modules within Asian and European laboratories”, Progress in Photovoltaics 27, 603-622, 2019. doi:10.1002/pip.3134
  4. Fraunhofer ISE, “CalLab PV Modules (accredited calibration laboratory), web page accessed November 2026”, Fraunhofer ISE, 2026. Source
  5. K. Maham et al., “Spectral mismatch uncertainty estimation in solar cell calibration using Monte Carlo simulation”, IEEE Journal of Photovoltaics 13, 899-904, 2023. doi:10.1109/JPHOTOV.2023.3311890
  6. IEC, “IEC 60904-9:2020, Photovoltaic devices, Part 9: Classification of solar simulator characteristics (edition 3.0)”, International Electrotechnical Commission, 2020. Source
  7. E. Salis, D. Pavanello, M. Field et al., “Improvements in world-wide intercomparison of PV module calibration”, Solar Energy 155, 1451-1461, 2017. doi:10.1016/j.solener.2017.07.081
  8. D. Chojniak, A. Schmid, J. Hohl-Ebinger, G. Siefer, S. W. Glunz, “Traceable efficiency determination of large-area perovskite on silicon tandem modules: insights from a calibration laboratory's perspective”, Progress in Photovoltaics 33, 815-827, 2025. doi:10.1002/pip.3928
  9. H. Müllejans, W. Zaaiman, R. Galleano, “Analysis and mitigation of measurement uncertainties in the traceability chain for the calibration of photovoltaic devices”, Measurement Science and Technology 20, 075101, 2009. doi:10.1088/0957-0233/20/7/075101
Related pages
Measurement uncertainty and traceabilityGuide: reading an uncertainty budgetGlossary: repeatability vs uncertainty

How this page is written: claims about physics and standards cite published sources, claims about Pasan equipment repeat only what the product pages state, and no other manufacturer is named. Editorial method

Rather we contact you?

Leave your email and we will get back to you. No form to fill in.

We only use your address to answer you. Privacy policy