The measurement problem
Buyers sometimes ask for an "ISO 17025 certified simulator". Accreditation applies to a laboratory's competence for defined measurements within a scope. What matters is what accreditation demands from the simulator, and what it leaves unproven.
The physics behind it
- 6.4 Equipment: capable of the required accuracy, calibrated when it affects the validity of results [1].
- 6.5 Metrological traceability: an unbroken chain to the SI; for PV through IEC 60904-4 [2].
- 7.2 Selection, verification and validation of methods.
- 7.6 Evaluation of measurement uncertainty, following the GUM.
- 7.7 Ensuring the validity of results, including proficiency testing or interlaboratory comparison; PV module intercomparisons use the En number [3].
What the measurement must achieve
- Classification of the simulator per IEC 60904-9 at the operating conditions, re-checked periodically, since classification is not constant [4].
- Calibrated reference devices traceable via IEC 60904-2 and -4.
- A calibrated spectroradiometer for the simulator spectrum used in mismatch correction.
- Voltage and current channels of the electronic load and the temperature sensors calibratable against external traceable references.
- A full uncertainty budget per measurand and technology, and participation in intercomparisons.
Applicable standards
| Standard | What it governs here |
|---|---|
| ISO/IEC 17025:2017General requirements for the competence of testing and calibration laboratories | The accreditation standard (third edition, 2017). |
| JCGM 100:2008 (GUM)Evaluation of measurement data: Guide to the expression of uncertainty in measurement | Uncertainty evaluation. |
| IEC 60904-4:2019Procedures for establishing calibration traceability of PV reference devices | Traceability of PV reference devices. |
| IEC 60904-9:2020 (edition 3)Classification of solar simulator characteristics | Simulator classification. |
| IEC 60904-8:2014Measurement of spectral responsivity of a PV device | Spectral responsivity measurement, with ISO/IEC 17025 requirements. |
Measurement technologies: advantages and limitations
Accredited laboratories use pulsed xenon, steady-state and LED systems: the standard is technology neutral. Accreditation does not certify a simulator model or class, does not extend to measurements outside the scope, and does not guarantee agreement beyond stated uncertainties: in a round robin of accredited laboratories on high-efficiency modules the largest Pmax deviation was 1.59 % [5].
Where Pasan equipment fits
- WPVS transimpedance amplifierReads non-shunted WPVS reference cells, so an accredited laboratory can calibrate every simulator against the same chain.
- HighLIGHT A++ ReferenceReference simulator with factory and on-site acceptance tests documenting the chain.
- Spectral Response KitTwo-channel spectroradiometer for IEC 60904-9 ed.3 classification and mismatch data.
Independent evidence
- Public sources name Pasan simulators at ISO/IEC 17025 accredited laboratories, for example JRC ESTI, SUPSI PVLab, Kiwa PVEL and USP-IEE: see the Scientific Validation database. The accreditation belongs to those laboratories, not to the equipment.
Technical questions
- Can a simulator be ISO/IEC 17025 certified?
- No. Laboratories are accredited for a scope; the simulator is equipment within it [1].
- What does accreditation require of the simulator?
- Calibration and traceability of the measurement chain, validated methods and an evaluated uncertainty.
- How do accredited laboratories prove ongoing competence?
- Through monitoring that includes proficiency tests and intercomparisons (clause 7.7); PV intercomparisons use En numbers [3].
- Does an accredited result guarantee a specific accuracy?
- It guarantees a stated, evaluated uncertainty, for example ±1.1 % for monofacial module Pmpp at Fraunhofer ISE CalLab [6].
References
- , “ISO/IEC 17025:2017, General requirements for the competence of testing and calibration laboratories”, International Organization for Standardization, 2017. Source
- , “IEC 60904-4:2019, Photovoltaic reference devices: procedures for establishing calibration traceability (edition 2.0)”, International Electrotechnical Commission, 2019. Source
- , “Improvements in world-wide intercomparison of PV module calibration”, Solar Energy 155, 1451-1461, 2017. doi:10.1016/j.solener.2017.07.081
- , “IEC 60904-9:2020, Photovoltaic devices, Part 9: Classification of solar simulator characteristics (edition 3.0)”, International Electrotechnical Commission, 2020. Source
- , “Electrical characterization intercomparison of high-efficiency c-Si modules within Asian and European laboratories”, Progress in Photovoltaics 27, 603-622, 2019. doi:10.1002/pip.3134
- , “CalLab PV Modules (accredited calibration laboratory), web page accessed November 2026”, Fraunhofer ISE, 2026. Source
How this page is written: claims about physics and standards cite published sources, claims about Pasan equipment repeat only what the product pages state, and no other manufacturer is named. Editorial method