What ISO/IEC 17025 requires of a PV testing or calibration laboratory, and of its solar simulator

ISO/IEC 17025:2017 accredits a laboratory for a defined scope, never an instrument: for the solar simulator it requires calibrated, traceable equipment (clauses 6.4 and 6.5), validated methods (7.2), an evaluated measurement uncertainty (7.6) and ongoing checks such as proficiency tests (7.7) [1], so the useful questions to a supplier are whether every part of the measurement chain can be calibrated externally and documented.

The measurement problem

Buyers sometimes ask for an "ISO 17025 certified simulator". Accreditation applies to a laboratory's competence for defined measurements within a scope. What matters is what accreditation demands from the simulator, and what it leaves unproven.

The physics behind it

  • 6.4 Equipment: capable of the required accuracy, calibrated when it affects the validity of results [1].
  • 6.5 Metrological traceability: an unbroken chain to the SI; for PV through IEC 60904-4 [2].
  • 7.2 Selection, verification and validation of methods.
  • 7.6 Evaluation of measurement uncertainty, following the GUM.
  • 7.7 Ensuring the validity of results, including proficiency testing or interlaboratory comparison; PV module intercomparisons use the En number [3].

What the measurement must achieve

  • Classification of the simulator per IEC 60904-9 at the operating conditions, re-checked periodically, since classification is not constant [4].
  • Calibrated reference devices traceable via IEC 60904-2 and -4.
  • A calibrated spectroradiometer for the simulator spectrum used in mismatch correction.
  • Voltage and current channels of the electronic load and the temperature sensors calibratable against external traceable references.
  • A full uncertainty budget per measurand and technology, and participation in intercomparisons.

Applicable standards

Applicable standards
StandardWhat it governs here
ISO/IEC 17025:2017General requirements for the competence of testing and calibration laboratoriesThe accreditation standard (third edition, 2017).
JCGM 100:2008 (GUM)Evaluation of measurement data: Guide to the expression of uncertainty in measurementUncertainty evaluation.
IEC 60904-4:2019Procedures for establishing calibration traceability of PV reference devicesTraceability of PV reference devices.
IEC 60904-9:2020 (edition 3)Classification of solar simulator characteristicsSimulator classification.
IEC 60904-8:2014Measurement of spectral responsivity of a PV deviceSpectral responsivity measurement, with ISO/IEC 17025 requirements.

Measurement technologies: advantages and limitations

Accredited laboratories use pulsed xenon, steady-state and LED systems: the standard is technology neutral. Accreditation does not certify a simulator model or class, does not extend to measurements outside the scope, and does not guarantee agreement beyond stated uncertainties: in a round robin of accredited laboratories on high-efficiency modules the largest Pmax deviation was 1.59 % [5].

Where Pasan equipment fits

Independent evidence

  • Public sources name Pasan simulators at ISO/IEC 17025 accredited laboratories, for example JRC ESTI, SUPSI PVLab, Kiwa PVEL and USP-IEE: see the Scientific Validation database. The accreditation belongs to those laboratories, not to the equipment.

Technical questions

Can a simulator be ISO/IEC 17025 certified?
No. Laboratories are accredited for a scope; the simulator is equipment within it [1].
What does accreditation require of the simulator?
Calibration and traceability of the measurement chain, validated methods and an evaluated uncertainty.
How do accredited laboratories prove ongoing competence?
Through monitoring that includes proficiency tests and intercomparisons (clause 7.7); PV intercomparisons use En numbers [3].
Does an accredited result guarantee a specific accuracy?
It guarantees a stated, evaluated uncertainty, for example ±1.1 % for monofacial module Pmpp at Fraunhofer ISE CalLab [6].

References

  1. ISO/IEC, “ISO/IEC 17025:2017, General requirements for the competence of testing and calibration laboratories”, International Organization for Standardization, 2017. Source
  2. IEC, “IEC 60904-4:2019, Photovoltaic reference devices: procedures for establishing calibration traceability (edition 2.0)”, International Electrotechnical Commission, 2019. Source
  3. E. Salis, D. Pavanello, M. Field et al., “Improvements in world-wide intercomparison of PV module calibration”, Solar Energy 155, 1451-1461, 2017. doi:10.1016/j.solener.2017.07.081
  4. IEC, “IEC 60904-9:2020, Photovoltaic devices, Part 9: Classification of solar simulator characteristics (edition 3.0)”, International Electrotechnical Commission, 2020. Source
  5. C. Monokroussos et al., “Electrical characterization intercomparison of high-efficiency c-Si modules within Asian and European laboratories”, Progress in Photovoltaics 27, 603-622, 2019. doi:10.1002/pip.3134
  6. Fraunhofer ISE, “CalLab PV Modules (accredited calibration laboratory), web page accessed November 2026”, Fraunhofer ISE, 2026. Source
Related pages
Guide: ISO/IEC 17025 and your simulatorReference module calibrationArticle: ISO/IEC 17025 accreditation for module measurement

How this page is written: claims about physics and standards cite published sources, claims about Pasan equipment repeat only what the product pages state, and no other manufacturer is named. Editorial method

Rather we contact you?

Leave your email and we will get back to you. No form to fill in.

We only use your address to answer you. Privacy policy