The measurement problem
A simulator measures irradiance only through a reference device. Each step of the chain from the SI to the working reference adds uncertainty, and in a module budget the reference calibration is often the largest single term [4].
The physics behind it
A reference cell's short-circuit current scales with effective irradiance weighted by its spectral responsivity; its calibration therefore needs its spectral responsivity, linearity and temperature coefficient and the spectrum of the calibration source [5]. The World Photovoltaic Scale (WPVS) is the international programme that made primary reference cells comparable through round-robin calibration traceable to the SI [6].
| Level | Value | Source |
|---|---|---|
| Primary WPVS cell, PTB laser DSR | 0.4 % (Isc at STC) | [2] |
| JRC ESTI reference cell set (multiple primary calibrations) | 0.23 % | [1] |
| Primary methods compared (PTB, NLR, AIST, 2026) | All within uncertainty; maximum deviations 0.27 to 0.59 % | [7] |
| Module Pmax, JRC ESTI chain (2009) | From ±2.6 % to below ±2 % using the cell set | [4] |
| Module Pmax, accredited, Fraunhofer ISE CalLab (web page) | ±1.1 % monofacial, ±1.4 % bifacial (coverage not stated) | [8] |
What the measurement must achieve
- Reference devices built, characterised and documented per IEC 60904-2:2023, which added calibration at maximum power and explicit traceability requirements [5].
- An unbroken, documented chain to the SI per IEC 60904-4:2019, which added the working reference to the chain [9].
- A matched reference (same technology) to minimise spectral mismatch and non-uniformity effects, or a full mismatch correction.
- For production reference modules: as identical as possible to the product, stabilised before first calibration (for example 5 kWh/m² light soaking), checked after dark storage, two per design with overlapping calibration intervals, traceability checked yearly [3].
- Calibration on Pmax compensates simulator non-uniformity better than calibration on Isc [3].
Applicable standards
| Standard | What it governs here |
|---|---|
| IEC 60904-2:2023Requirements for photovoltaic reference devices | Requirements for reference devices, including calibration at maximum power (2023). |
| IEC 60904-4:2019Procedures for establishing calibration traceability of PV reference devices | Procedures for establishing traceability; working references in the chain (2019). |
| IEC 60904-7:2019Computation of the spectral mismatch correction | Mismatch correction when reference and device differ. |
| IEC 60904-8:2014Measurement of spectral responsivity of a PV device | Spectral responsivity measurement, with ISO/IEC 17025 requirements since 2014. |
| IEC 60904-10:2020Methods of linear dependence and linearity measurements | Linearity of the reference device. |
| ISO/IEC 17025:2017General requirements for the competence of testing and calibration laboratories | Traceability and uncertainty requirements of calibration laboratories. |
Measurement technologies: advantages and limitations
Calibration with simulated sunlight needs A+-level uniformity and an accurately known spectrum; pulsed and steady-state simulators are both used. LED simulators used for calibration need traceable calibration of each LED channel [10]. No source supports a general claim that one lamp technology yields lower calibration uncertainty: the reference chain and the mismatch dominate [4].
Where Pasan equipment fits
- HighLIGHT A++ ReferencePasan's reference simulator (spectral match ±6.25 %, non-uniformity and long-term instability below 0.5 %; A++ is Pasan's convention for half of the IEC A+ limits), traceable to METAS.
- WPVS transimpedance amplifierConverts the current of non-shunted WPVS reference cells into a voltage with ±0.1 % accuracy at 25 °C.
- Spectral Response KitDevice spectral response and simulator spectrum for mismatch-corrected calibration.
Independent evidence
- Public laboratory documents name Pasan simulators at reference and accredited laboratories, among them JRC ESTI and Fraunhofer ISE CalLab: see the Scientific Validation database. Using a simulator to calibrate secondary references does not make the result a primary calibration.
Technical questions
- What is the WPVS?
- The World Photovoltaic Scale, an international reference cell calibration programme traceable to the SI through round-robin calibration of primary reference cells [6].
- What is the lowest published uncertainty for a primary reference cell?
- 0.4 % (k = 2) for WPVS cells with PTB's laser DSR [2]; a multiply calibrated set at JRC ESTI reached 0.23 % (k = 2) [1].
- Do different primary methods agree?
- Yes: in a 2026 PTB, NLR and AIST comparison all results agreed within their uncertainties [7].
- How often should a golden module be recalibrated?
- The JRC industry guidelines recommend yearly traceability checks and overlapping pairs of reference modules [3].
References
- , “Reduction of uncertainties for photovoltaic reference cells”, Metrologia 52, 646-653, 2015. doi:10.1088/0026-1394/52/5/646
- , “Design, realization and uncertainty analysis of a laser-based primary calibration facility for solar cells at PTB”, Measurement 51, 457-463, 2014. doi:10.1016/j.measurement.2013.12.001
- , “Guidelines for PV power measurement in industry”, EUR 24359 EN, Publications Office of the European Union, 2010. Source
- , “Analysis and mitigation of measurement uncertainties in the traceability chain for the calibration of photovoltaic devices”, Measurement Science and Technology 20, 075101, 2009. doi:10.1088/0957-0233/20/7/075101
- , “IEC 60904-2:2023, Photovoltaic devices, Part 2: Requirements for photovoltaic reference devices (edition 4.0)”, International Electrotechnical Commission, 2023. Source
- , “The world photovoltaic scale: an international reference cell calibration program”, Progress in Photovoltaics 7, 287-297, 1999. doi:10.1002/(SICI)1099-159X(199907/08)7:4<287::AID-PIP259>3.0.CO;2-I
- , “International intercomparison of primary photovoltaic reference cell calibrations with excellent agreement”, Solar RRL 10, 2026. doi:10.1002/solr.70475
- , “CalLab PV Modules (accredited calibration laboratory), web page accessed November 2026”, Fraunhofer ISE, 2026. Source
- , “IEC 60904-4:2019, Photovoltaic reference devices: procedures for establishing calibration traceability (edition 2.0)”, International Electrotechnical Commission, 2019. Source
- , “Traceable efficiency determination of large-area perovskite on silicon tandem modules: insights from a calibration laboratory's perspective”, Progress in Photovoltaics 33, 815-827, 2025. doi:10.1002/pip.3928
How this page is written: claims about physics and standards cite published sources, claims about Pasan equipment repeat only what the product pages state, and no other manufacturer is named. Editorial method