The measurement problem
Cells in a series string under unequal irradiance produce unequal currents; the module current is pulled toward the weakest cells, distorting Isc and fill factor. The reference device adds a second error if it sees a different irradiance than the module average.
The physics behind it
Non-uniformity mainly lowers Isc and raises fill factor; in one JRC example, Pmax was not affected below 5 % non-uniformity [2]. The effect depends on the spatial pattern relative to cells and bypass diodes, the cells' shunt resistance and the reference position; a reference module calibrated on Pmax compensates non-uniformity better than one calibrated on Isc [2]. In large modules, non-uniformity is a leading term of the uncertainty budget at reference laboratories.
What the measurement must achieve
- Classify non-uniformity over the designated test area per IEC 60904-9:2020 clause 5.3, at each irradiance level used.
- Re-measure periodically and after lamp exchange, power changes, inserts (filters, masks) or room changes: the classification is not constant [1].
- Use a reference of the module type, ideally calibrated on Pmax, to compensate residual non-uniformity [2].
- For bifacial single-side methods, keep class B or better at the elevated irradiance [3].
Applicable standards
| Standard | What it governs here |
|---|---|
| IEC 60904-9:2020 (edition 3)Classification of solar simulator characteristics | Definition, limits and procedure for spatial non-uniformity. |
| IEC 60904-1:2020Measurement of photovoltaic current-voltage characteristics | Annex D: Monte Carlo example of non-uniformity effects on a module. |
| IEC 61215-2:2021Terrestrial PV modules, design qualification and type approval: test procedures | MQT 02 allows class B uniformity (CBA) for very large modules. |
| IEC TS 60904-1-2:2024Measurement of current-voltage characteristics of bifacial PV devices | Class B or better at elevated irradiance for bifacial measurement. |
Measurement technologies: advantages and limitations
Large-area pulsed xenon with a long throw tends to give smooth, low-spatial-frequency patterns; close-proximity multi-source LED or reflector systems have patterns set by the emitter grid and optics. No source supports a universal claim that one lamp type is more uniform: uniformity is a property of the optical design and is measured per instrument [1].
Where Pasan equipment fits
- HighLIGHT A++ ReferencePasan states non-uniformity below 0.5 % (half of the IEC A+ limit; A++ is Pasan's convention, not an IEC class).
- HighLIGHT 100ms LabNon-uniformity below 1 % (A+) over a test area up to 2.8 × 1.6 m.
Technical questions
- Does class A+ uniformity guarantee less than 1 % error?
- No. The class is a max/min metric; the error depends on the pattern and the device, and the classification does not predict power uncertainty [1].
- What does non-uniformity do to the I-V curve?
- Isc decreases and fill factor increases [2].
- Is a reference module calibrated on Isc or on Pmax better?
- Pmax-calibrated references compensate non-uniformity better [2].
- How often should uniformity be re-measured?
- Periodically, and after lamp exchange, power changes, inserts or room changes [1].
References
- , “IEC 60904-9:2020, Photovoltaic devices, Part 9: Classification of solar simulator characteristics (edition 3.0)”, International Electrotechnical Commission, 2020. Source
- , “Guidelines for PV power measurement in industry”, EUR 24359 EN, Publications Office of the European Union, 2010. Source
- , “IEC TS 60904-1-2:2024+AMD1:2026, Measurement of current-voltage characteristics of bifacial photovoltaic devices”, International Electrotechnical Commission, 2024. Source