Measuring perovskite-silicon and other tandem devices

In a two-terminal tandem the subcells are in series, so the current is set by whichever junction limits under the test spectrum; the simulator spectrum must therefore be adjusted, using the spectral responsivity of each junction (IEC 60904-8-1) and one reference device per junction, until each subcell generates its AM1.5G current (IEC 60904-1-1), and the perovskite top cell then needs a steady-state power measurement [1][2].

The measurement problem

A spectrum slightly bluer or redder than AM1.5G changes which subcell limits the current, the fill factor and even the hysteresis of a perovskite-silicon tandem [2]. A tandem efficiency is therefore meaningless without the spectral adjustment behind it.

The physics behind it

  • Current matching: the I-V shape near the matching point is sensitive to small spectral changes [2].
  • A slight change of current around the matching point changes hysteresis markedly, because the perovskite subcell is biased differently depending on which subcell limits [2].
  • Mobile ions in the perovskite absorber add the slow transients of single-junction perovskite devices.

What the measurement must achieve

  • Measure the spectral responsivity of each junction (IEC 60904-8-1).
  • Adjust the simulator so that each subcell generates the current it would under AM1.5G, with one reference device (isotype or filtered component cell) per junction (IEC 60904-1-1).
  • Compute a mismatch factor per junction (IEC 60904-7): Fraunhofer ISE readjusted to within 0.1 % of unity [1].
  • Determine the steady-state power: on a stable 1.682 m² module, MPP tracking, asymptotic Pmax and conventional sweeps agreed within 0.05 % [1].
  • Report a spectral sensitivity analysis from top-limited to bottom-limited conditions.

Applicable standards

Applicable standards
StandardWhat it governs here
IEC 60904-1-1:2017Measurement of current-voltage characteristics of multi-junction PV devicesI-V of multi-junction devices with spectral adjustment per junction.
IEC 60904-8-1:2017Measurement of spectral responsivity of multi-junction PV devicesSpectral responsivity of each junction, the prerequisite of the adjustment.
IEC 60904-7:2019Computation of the spectral mismatch correctionMismatch factors, extended to each junction.
IEC TR 63228:2019Measurement protocols for PV devices based on organic, dye-sensitized or perovskite materialsPerovskite-specific measurement challenges.
IEC 60904-9:2020 (edition 3)Classification of solar simulator characteristicsClassification of the simulator.

Measurement technologies: advantages and limitations

A single-source pulsed xenon simulator has one fixed spectrum and cannot set two junction currents independently; per-junction adjustment needs filters or additional sources, and its pulses are short compared with perovskite transients. Multi-source simulators (xenon with halogen, or multi-channel LED) are the standard approach [3]. LED tunability enables adjustment and spectral sweeps, but each LED channel needs traceable calibration and a warm-up for spectral stability (90 min in Fraunhofer's procedure, with up to 1.5 % adjustment uncertainty per subcell found in a spectral-variation study) [1]. Industrial in-line concepts approximate EQE from LED measurements, then measure I-V with an adjusted LED spectrum after a short preconditioning [4].

Where Pasan equipment fits

Independent evidence

  • A 1.682 m² perovskite-silicon tandem module was measured at 25.00 % ± 0.83 % (k = 2) by an accredited calibration laboratory, the mismatch contribution being about 0.2 % [1].

Technical questions

Why does a tandem efficiency depend on the spectrum?
The current-limiting subcell, the fill factor and even the hysteresis change with small spectral shifts [2].
What does IEC 60904-1-1 require first?
The spectral responsivity of each junction, measured per IEC 60904-8-1.
Can a single xenon flash measure a perovskite-silicon tandem?
Not correctly: it cannot adjust two junction currents independently, and its pulse is far shorter than the perovskite transients.
What uncertainty is achievable for a full-size tandem module?
25.00 % ± 0.83 % (k = 2) at an accredited laboratory in 2025 [1].

References

  1. D. Chojniak, A. Schmid, J. Hohl-Ebinger, G. Siefer, S. W. Glunz, “Traceable efficiency determination of large-area perovskite on silicon tandem modules: insights from a calibration laboratory's perspective”, Progress in Photovoltaics 33, 815-827, 2025. doi:10.1002/pip.3928
  2. C. Messmer et al., “Toward more reliable measurement procedures of perovskite-silicon tandem solar cells: the role of transient device effects and measurement conditions”, Progress in Photovoltaics 33, 126-142, 2024. doi:10.1002/pip.3782
  3. D. Chojniak et al., “A precise method for the spectral adjustment of LED and multi-light source solar simulators”, Progress in Photovoltaics 32, 372-389, 2024. doi:10.1002/pip.3776
  4. M. C. Schubert et al., “Industrial production of perovskite-silicon tandem solar cells: the characterization challenge”, Solar RRL 10, 2026. doi:10.1002/solr.202500766
Related pages
Perovskite cells and modulesSpectral mismatch correctionArticle: space cell testing, AM0, multi-junction

How this page is written: claims about physics and standards cite published sources, claims about Pasan equipment repeat only what the product pages state, and no other manufacturer is named. Editorial method

Rather we contact you?

Leave your email and we will get back to you. No form to fill in.

We only use your address to answer you. Privacy policy