The measurement problem
A spectrum slightly bluer or redder than AM1.5G changes which subcell limits the current, the fill factor and even the hysteresis of a perovskite-silicon tandem [2]. A tandem efficiency is therefore meaningless without the spectral adjustment behind it.
The physics behind it
- Current matching: the I-V shape near the matching point is sensitive to small spectral changes [2].
- A slight change of current around the matching point changes hysteresis markedly, because the perovskite subcell is biased differently depending on which subcell limits [2].
- Mobile ions in the perovskite absorber add the slow transients of single-junction perovskite devices.
What the measurement must achieve
- Measure the spectral responsivity of each junction (IEC 60904-8-1).
- Adjust the simulator so that each subcell generates the current it would under AM1.5G, with one reference device (isotype or filtered component cell) per junction (IEC 60904-1-1).
- Compute a mismatch factor per junction (IEC 60904-7): Fraunhofer ISE readjusted to within 0.1 % of unity [1].
- Determine the steady-state power: on a stable 1.682 m² module, MPP tracking, asymptotic Pmax and conventional sweeps agreed within 0.05 % [1].
- Report a spectral sensitivity analysis from top-limited to bottom-limited conditions.
Applicable standards
| Standard | What it governs here |
|---|---|
| IEC 60904-1-1:2017Measurement of current-voltage characteristics of multi-junction PV devices | I-V of multi-junction devices with spectral adjustment per junction. |
| IEC 60904-8-1:2017Measurement of spectral responsivity of multi-junction PV devices | Spectral responsivity of each junction, the prerequisite of the adjustment. |
| IEC 60904-7:2019Computation of the spectral mismatch correction | Mismatch factors, extended to each junction. |
| IEC TR 63228:2019Measurement protocols for PV devices based on organic, dye-sensitized or perovskite materials | Perovskite-specific measurement challenges. |
| IEC 60904-9:2020 (edition 3)Classification of solar simulator characteristics | Classification of the simulator. |
Measurement technologies: advantages and limitations
A single-source pulsed xenon simulator has one fixed spectrum and cannot set two junction currents independently; per-junction adjustment needs filters or additional sources, and its pulses are short compared with perovskite transients. Multi-source simulators (xenon with halogen, or multi-channel LED) are the standard approach [3]. LED tunability enables adjustment and spectral sweeps, but each LED channel needs traceable calibration and a warm-up for spectral stability (90 min in Fraunhofer's procedure, with up to 1.5 % adjustment uncertainty per subcell found in a spectral-variation study) [1]. Industrial in-line concepts approximate EQE from LED measurements, then measure I-V with an adjusted LED spectrum after a short preconditioning [4].
Where Pasan equipment fits
- Nexun PRO Max with the steady-state add-on (Avalon ST)Per-wavelength control, EQE and multi-junction capability, AM1.5G and AM0; with the steady-state add-on it is sold as the Nexun Perovskite for perovskite and perovskite-silicon tandem testing.
- HighLIGHT SAT multi-junctionLarge-area AM0 tester for 3J, 4J and 5J space cells and satellite panels.
- Guide (avalon.solar): how to measure a perovskite-silicon tandemStep-by-step procedure on a Nexun: EQE per junction, one reference per junction, band tuning, steady-state preconditioning, I-V and stabilised power.
Independent evidence
- A 1.682 m² perovskite-silicon tandem module was measured at 25.00 % ± 0.83 % (k = 2) by an accredited calibration laboratory, the mismatch contribution being about 0.2 % [1].
Technical questions
- Why does a tandem efficiency depend on the spectrum?
- The current-limiting subcell, the fill factor and even the hysteresis change with small spectral shifts [2].
- What does IEC 60904-1-1 require first?
- The spectral responsivity of each junction, measured per IEC 60904-8-1.
- Can a single xenon flash measure a perovskite-silicon tandem?
- Not correctly: it cannot adjust two junction currents independently, and its pulse is far shorter than the perovskite transients.
- What uncertainty is achievable for a full-size tandem module?
- 25.00 % ± 0.83 % (k = 2) at an accredited laboratory in 2025 [1].
References
- , “Traceable efficiency determination of large-area perovskite on silicon tandem modules: insights from a calibration laboratory's perspective”, Progress in Photovoltaics 33, 815-827, 2025. doi:10.1002/pip.3928
- , “Toward more reliable measurement procedures of perovskite-silicon tandem solar cells: the role of transient device effects and measurement conditions”, Progress in Photovoltaics 33, 126-142, 2024. doi:10.1002/pip.3782
- , “A precise method for the spectral adjustment of LED and multi-light source solar simulators”, Progress in Photovoltaics 32, 372-389, 2024. doi:10.1002/pip.3776
- , “Industrial production of perovskite-silicon tandem solar cells: the characterization challenge”, Solar RRL 10, 2026. doi:10.1002/solr.202500766
How this page is written: claims about physics and standards cite published sources, claims about Pasan equipment repeat only what the product pages state, and no other manufacturer is named. Editorial method