The measurement problem
A reference cell and a module with different spectral responsivities read different effective irradiances under any spectrum other than the reference spectrum. Without correction, the module's current and power carry the product of both deviations.
The physics behind it
With E_ref the AM1.5G spectrum of IEC 60904-3, E_meas the simulator spectrum in the test plane and s_ref, s_DUT the relative spectral responsivities, the factor is SMM = [∫E_ref s_ref dλ · ∫E_meas s_DUT dλ] / [∫E_meas s_ref dλ · ∫E_ref s_DUT dλ]. Absolute scales cancel, so relative responsivities suffice. IEC 60904-7:2019 renamed the factor SMM and stresses that it applies only to the combination of reference, device and spectrum for which it was computed [3][4].
What the measurement must achieve
- Spectral responsivity of the device (IEC 60904-8; a cell of the same bill of materials is accepted for MQT 02).
- Spectral responsivity of the reference device.
- Simulator spectrum in the test plane, measured with an adequate spectroradiometer and, for pulsed simulators, integrated over the acquisition time because spectral match can change during the pulse [5].
- A simulator spectrum covering the device's whole responsivity range, as IEC 61215-2 recommends.
- An uncertainty for the factor itself, including correlations.
Applicable standards
| Standard | What it governs here |
|---|---|
| IEC 60904-7:2019Computation of the spectral mismatch correction | Definition and application of the SMM. |
| IEC 60904-8:2014Measurement of spectral responsivity of a PV device | Spectral responsivity of the device. |
| IEC 60904-3:2019Measurement principles and reference spectral irradiance data (AM1.5G) | AM1.5G reference spectrum. |
| IEC 60904-9:2020 (edition 3)Classification of solar simulator characteristics | Spectral match classification; Annex A sensitivity analysis of mismatch error. |
| IEC 61215-2:2021Terrestrial PV modules, design qualification and type approval: test procedures | MQT 02: mismatch correction or a matched reference device. |
Measurement technologies: advantages and limitations
Spectral class is assessed in six bands only; spectral deviation (SPD) captures deviations inside the bands that the class does not see [5]. Xenon spectra have line structure in the near infrared that must be filtered; LED spectra are sums of discrete emitters that can leave gaps unless broadband channels fill them, and per-channel calibration adds its own uncertainty [6]. A matched reference device (same technology) reduces the correction to near unity, which is why production uses reference modules of the product type [7].
| Case | Value | Source |
|---|---|---|
| c-Si module vs c-Si reference on class AAA simulators | Correction reduced error by about 2 % | [1] |
| Amorphous silicon p-i-n | Voltage-dependent mismatch of 1 to 4 % in Pmpp | [8] |
| CdTe, two silicon reference choices | Isc differed by 1.1 % on average | [9] |
| Perovskite-silicon tandem module, per junction | Factors within 0.1 % of unity after adjustment; 0.2 % contribution | [6] |
| Uncertainty of the factor (cell calibration) | 0.06 % to 1.26 % (k = 2) | [2] |
Where Pasan equipment fits
- Spectral Response KitAbout 20 bandpass filters (50 nm FWHM, 300 to 1200 nm) for device spectral response under the real flash, and a two-channel spectroradiometer for the simulator spectrum and IEC 60904-9 ed.3 classification.
- HighLIGHT 100ms Lab15-filter module spectral response option (400 to 1100 nm).
Technical questions
- If my simulator is class A+ spectrally, can I skip the correction?
- No. IEC 60904-9 itself states that the classification does not predict power measurement uncertainty [5], and corrections of about 2 % were found even for silicon on AAA simulators [1].
- Do I need absolute spectral responsivity?
- No, relative responsivities suffice because the scale factors cancel.
- Is the factor transferable between simulators?
- No, it is valid only for that combination of reference, device and spectrum.
- Why integrate the spectrum over the acquisition time on a flasher?
- Because spectral match can change during the pulse [5].
References
- , “Electrical characterization intercomparison of high-efficiency c-Si modules within Asian and European laboratories”, Progress in Photovoltaics 27, 603-622, 2019. doi:10.1002/pip.3134
- , “Spectral mismatch uncertainty estimation in solar cell calibration using Monte Carlo simulation”, IEEE Journal of Photovoltaics 13, 899-904, 2023. doi:10.1109/JPHOTOV.2023.3311890
- , “Uncertainty of the spectral mismatch correction factor in STC measurements on photovoltaic devices”, Progress in Photovoltaics 19, 573-579, 2011. doi:10.1002/pip.1059
- , “An uncertainty analysis of the spectral correction factor”, 23rd IEEE Photovoltaic Specialists Conference, 1993. doi:10.1109/PVSC.1993.346955
- , “IEC 60904-9:2020, Photovoltaic devices, Part 9: Classification of solar simulator characteristics (edition 3.0)”, International Electrotechnical Commission, 2020. Source
- , “Traceable efficiency determination of large-area perovskite on silicon tandem modules: insights from a calibration laboratory's perspective”, Progress in Photovoltaics 33, 815-827, 2025. doi:10.1002/pip.3928
- , “Guidelines for PV power measurement in industry”, EUR 24359 EN, Publications Office of the European Union, 2010. Source
- , “Effects of spectrum on the power rating of amorphous silicon photovoltaic devices”, Progress in Photovoltaics, 2011. doi:10.1002/pip.1080
- , “The components of variance for calibration methods of thin film CdTe photovoltaic modules”, 39th IEEE Photovoltaic Specialists Conference, 2013. doi:10.1109/PVSC.2013.6744232
How this page is written: claims about physics and standards cite published sources, claims about Pasan equipment repeat only what the product pages state, and no other manufacturer is named. Editorial method