The measurement problem
Buyers often ask which simulator is best for CdTe. The honest answer is that most of the error on a CdTe measurement comes from the module's state and from spectral mismatch, not from the lamp [2][3]. A simulator suits CdTe when it makes those two terms small and documented.
The physics behind it
CdTe and CdSeTe absorbers collect light from about 300 to 900 nm [1][4]. The infrared part of a simulator spectrum, where xenon has strong emission lines, reaches a silicon reference cell but not the module, so a spectrum that is class A+ for silicon can still produce a mismatch of a few percent for CdTe if it is not corrected [5].
CdTe modules are metastable: their power depends on recent light, bias and temperature [6]. They are, on the other hand, low-capacitance devices whose measured power does not depend on sweep time above about 1 ms [7].
What the measurement must achieve
- Spectral irradiance of the simulator measured at the test plane over 300 to 1200 nm, recorded with the calibration, for the IEC 60904-7 correction.
- Module-level spectral responsivity of the device type (IEC 60904-8), measured or supplied.
- A reference device matched to CdTe (a calibrated CdTe module of the same product) or a silicon reference with documented correction [8].
- An electronic load rated for the module's open-circuit voltage, with externally calibratable voltage and current channels.
- Module temperature measurement and control, with temperature coefficients of the product for correction.
- A reproducible preconditioning step: an integrated current-injection sequence with a fixed delay to the flash, validated against light soaking, or access to a light soaker [9].
Applicable standards
| Standard | What it governs here |
|---|---|
| IEC 60904-9:2020 (edition 3)Classification of solar simulator characteristics | Simulator classification; for CdTe the measured spectrum matters more than the class letter, because mismatch is corrected with it. |
| IEC 60904-7:2019Computation of the spectral mismatch correction | The mismatch correction itself. |
| IEC 60904-8:2014Measurement of spectral responsivity of a PV device | Spectral responsivity of the device. |
| IEC 61215-1-2:2021Special requirements for thin-film cadmium telluride (CdTe) modules | CdTe stabilisation before power measurement. |
| IEC 60904-2:2023Requirements for photovoltaic reference devices | Reference device requirements. |
Measurement technologies: advantages and limitations
Short-pulse and long-pulse xenon, steady-state and LED simulators can all measure a stabilised CdTe module; reference laboratories have done it with pulses of a few milliseconds [10]. LED spectra can be tuned closer to AM1.5G in the CdTe band, which modelling predicts would lower mismatch [11]; xenon spectra are smooth over the CdTe band but need correction for their infrared lines. Steady-state light can precondition and measure in one place, at the cost of heating the module during the measurement. None of these choices removes the need for a controlled preconditioning step.
Where Pasan equipment fits
- HighLIGHT 100ms LabMeasures CdTe modules after current-injection preconditioning, with the flash a fraction of a second after the injection; 15-filter set for module spectral response.
- Spectral Response KitBandpass filters for device spectral response (IEC 60904-8) and a spectroradiometer for the simulator spectrum: the two inputs of the mismatch correction.
- LED Light SoakerRegulated continuous light with load bias for MQT 19 stabilisation and validation of a current-injection recipe.
Technical questions
- Is an LED or a xenon simulator better for CdTe?
- Either can measure a stabilised CdTe module. What matters is a measured spectrum for the mismatch correction, a matched reference and controlled preconditioning. LED spectra can be tuned to reduce mismatch, which has so far been shown by modelling [11].
- Does a CdTe simulator need a long pulse?
- Not for the I-V curve: CdTe modules show no sweep-time effect above about 1 ms [7]. A long pulse matters for capacitive silicon such as HJT and TOPCon.
- Is A+ spectral class enough for CdTe?
- The class is graded against the silicon range up to 1200 nm. For CdTe, measure the spectrum and correct the mismatch with IEC 60904-7; the residual uncertainty of that correction is what enters the budget.
References
- , “Characterisation of performance of thin-film photovoltaic technologies”, IEA PVPS Task 13, Report IEA-PVPS T13-02:2014, 2014. Source
- , “Improvements in world-wide intercomparison of PV module calibration”, Solar Energy 155, 1451-1461, 2017. doi:10.1016/j.solener.2017.07.081
- , “Uncertainty in PV module measurement, part I: calibration of crystalline and thin-film modules”, IEEE Journal of Photovoltaics 3(3), 1016-1026, 2013. doi:10.1109/JPHOTOV.2013.2260595
- , “Recombination and bandgap engineering in CdSeTe/CdTe solar cells”, APL Materials 7, 071112, 2019. doi:10.1063/1.5098459
- , “Solar cell characterization”, Semiconductor Materials for Solar Photovoltaic Cells, Springer Series in Materials Science, 2016. doi:10.1007/978-3-319-20331-7_8
- , “Performance stabilization of CdTe PV modules using bias and light”, IEEE Journal of Photovoltaics 5(1), 344-349, 2015. doi:10.1109/JPHOTOV.2014.2370252
- , “Sweep time, spectral mismatch and light soaking in thin film module measurement”, 33rd IEEE Photovoltaic Specialists Conference, 2008. doi:10.1109/PVSC.2008.4922583
- , “The components of variance for calibration methods of thin film CdTe photovoltaic modules”, 39th IEEE Photovoltaic Specialists Conference, 2013. doi:10.1109/PVSC.2013.6744232
- , “Robust measurement of thin-film photovoltaic modules exhibiting light-induced transients”, Proc. SPIE 9561, 956108, 2015. doi:10.1117/12.2188827
- , “Preconditioning of thin-film PV modules through controlled light-soaking”, NREL PV Module Reliability Workshop, 2012. Source
- , “Optimization of LED-based solar simulators for cadmium telluride and microcrystalline silicon solar cells”, Energies 16, 5741, 2023. doi:10.3390/en16155741
How this page is written: claims about physics and standards cite published sources, claims about Pasan equipment repeat only what the product pages state, and no other manufacturer is named. Editorial method