- Institution
- Oerlikon Solar (text: 'With installation at Oerlikon's site'; site not specified); author affiliations Oerlikon Solar-Lab SA (Neuchâtel), Oerlikon Solar AG (Trübbach), Photovoltaic Consultant (Freiburg), Switzerland (author affiliations)
- Equipment named
- commercial solar simulator SunSim IIIb (Pasan SA, CH), modified for tunable spectrum with a longitudinal sliding test bed
- Application
- Spectral balancing by lamp intensity and distance, with PTB-calibrated top and bottom component reference cells, for traceable STC power measurement of Micromorph tandem modules
- Use
- Research
- Published uncertainty
- Component reference cells calibrated by PTB with uncertainty below 0.9% (k=2)
- Technologies
- Thin film, Multi-junction
- Topics
- Venue
- Conference paper, 26th European Photovoltaic Solar Energy Conference and Exhibition (EU PVSEC 2011), pp. 2744-2749, 2011
Significance
Establishes a traceable STC set-up for Micromorph tandem modules on a modified Pasan flasher, with 0.07% and 0.15% deviation from STC on the two component cells.
From the source
“Our technical approach takes the commercial solar simulator SunSim IIIb (Pasan SA, CH) as a basis.”
This entry summarises a third-party publication. It does not imply that the institution or the authors endorse Pasan SA or its products.




