Knowledge Center · Scientific validation · Topics
Amorphous and micromorph silicon: publications naming Pasan equipment
Amorphous silicon and micromorph (amorphous on microcrystalline) tandems degrade under light (Staebler-Wronski) and, as tandems, need spectral mismatch handled per junction.
6 publications · 6 institutions
- Secondary calibration2013TEL Solar metrology laboratory ('the company's metrology laboratory'); author affiliations TEL Solar-Lab SA (Neuchâtel), TEL Solar AG (Trübbach), Photovoltaic Consultant (Freiburg)Switzerland (author affiliations)
Power Measurement Analysis of Micromorph Tandem Modules
Equipment named: modified solar simulator Pasan SunSim IIIb; Pasan electrical load with 16 bit data acquisition. Application: STC power measurement and calibration of Micromorph reference modules, a 'fading' multi-flash procedure for transient effects, and spectrometric assessment of tandem current matching
- Accredited testing2010Photovoltaik Institut Berlin (PI Berlin)Germany
Actual issues on power measurement of photovoltaic modules
Equipment named: Pasan SSIIIb (class A, 3.0 m x 3.0 m). Application: Reference simulator at PI Berlin in a study of uniformity, spectral mismatch, tandem current mismatch and sweep-time (transient) effects on Pmax of crystalline and thin-film modules (a-Si, micromorph tandem, CIGS).
- Research2026Not stated for the flasher (thesis, University of Vienna; supervisor at AIT Austrian Institute of Technology)Austria
Short-Term Metastabilities and Power Rating in Thin-Film Photovoltaic Technologies
Equipment named: Pasan BV 66 active load and Pasan software for pulsed-light I-V measurement. Application: I-V measurement of amorphous silicon and micromorph modules during light-soaking pretreatment, and temperature correction of the curves, in a study of thin-film metastability and power rating
- Research2012Not stated; author affiliations Pasan SA (Neuchâtel), Oerlikon Solar AG (Trübbach), Oerlikon Solar-Lab SA (Neuchâtel), Photovoltaic Consultant (Freiburg)Switzerland (author affiliations)
Tandem Solar Simulator for Power Measurement of Micromorph Tandem Modules
Equipment named: Pasan High LIGHT electronic load; large-area tandem solar simulator with two independent, synchronized xenon flash lamps ('Blue' and 'Red'). Application: Development and qualification of a spectrally adjustable (±10% 'spectral swing') tandem simulator for Micromorph module power measurement at tandem STC, using top and bottom component cells
- Research2011Oerlikon Solar (text: 'With installation at Oerlikon's site'; site not specified); author affiliations Oerlikon Solar-Lab SA (Neuchâtel), Oerlikon Solar AG (Trübbach), Photovoltaic Consultant (Freiburg)Switzerland (author affiliations)
Traceable Flasher-Based Power Measurement of Micromorph Tandem Modules
Equipment named: commercial solar simulator SunSim IIIb (Pasan SA, CH), modified for tunable spectrum with a longitudinal sliding test bed. Application: Spectral balancing by lamp intensity and distance, with PTB-calibrated top and bottom component reference cells, for traceable STC power measurement of Micromorph tandem modules
- Research2011Technische Universität Berlin (with PI Berlin)Germany
Spectral Mismatch Correction for Micromorph Tandem Modules
Equipment named: Pasan Flasher. Application: Spectral mismatch correction for micromorph tandem modules, compared with LED and steady-state simulators.
These entries summarise third-party publications. They do not imply that the institutions or the authors endorse Pasan SA or its products.