Measuring heterojunction (HJT, SHJ) cells and modules without capacitive error

Heterojunction cells store so much charge that an I-V sweep faster than the device can follow splits the forward and reverse curves and biases the maximum power, by more than 20 % in extreme cases of fast sweeps [1]; a valid HJT measurement must be shown equivalent to steady state (IEC 60904-1:2020), typically by forward and reverse sweeps that agree, using a long enough pulse, a multi-flash or shaped-voltage method, or steady-state light [2].

The measurement problem

Silicon heterojunction (SHJ, HJT) devices combine long carrier lifetimes with high open-circuit voltage. Both raise the device's effective capacitance, so the current lags when the voltage is swept quickly. The result is hysteresis: the curve measured from short circuit to open circuit differs from the one measured the other way, and neither equals the steady-state curve that the module delivers in the field [3].

A 2021 review concluded that even 100 ms simulators are no longer enough for accurate direct I-V measurement of some recent large-cell n-type and SHJ modules, and compared eight mitigation methods with their uncertainties [2].

The physics behind it

Two capacitances act: the junction capacitance, dominant in reverse and low forward bias, and the diffusion capacitance, which grows with stored minority carriers in forward bias near the maximum power point and open circuit [1]. The error grows with capacitance, cell area and series resistance and is worst for short flashes [3]; it scales with the hold time per voltage point and the number of points [4], and the optimal scan time depends on irradiance [5].

Sweep direction matters in a predictable way: forward and reverse sweeps err in opposite directions, which is why their agreement is used as a diagnostic of quasi-steady state [1].

What the measurement must achieve

  • Demonstrate equivalence to steady-state performance (IEC 60904-1:2020 clause 4.4 and Annex B on capacitive devices).
  • Check forward and reverse sweeps, or two sweep durations, on each module type: they must agree within the target uncertainty.
  • Correct each I-V point with simultaneously measured irradiance; with multi-flash, the long-term instability spans the whole acquisition (IEC 60904-9:2020).
  • Control module temperature, particularly with long exposures.
  • Report the method, sweep time and direction with the result.

Applicable standards

Applicable standards
StandardWhat it governs here
IEC 60904-1:2020Measurement of photovoltaic current-voltage characteristicsEquivalence to steady state and Annex B (PV devices with capacitance): steady-state, quasi-steady-state and common methods.
IEC 60904-9:2020 (edition 3)Classification of solar simulator characteristicsSTI per data point, LTI over the I-V acquisition; with sectoring across flashes, LTI is the sum of the sector times.
IEC 61215-2:2021Terrestrial PV modules, design qualification and type approval: test proceduresMQT 02 maximum power determination in qualification.
IEC TS 63202-3:2023Measurement of current-voltage characteristics of bifacial photovoltaic cellsI-V of bifacial cells, most HJT cells being bifacial.

Measurement technologies: advantages and limitations

Measurement approaches for capacitive HJT devices
ApproachAdvantagesLimitations
Short single pulse (few ms), linear sweepFast, no heatingLarge capacitive error possible on HJT [1]; only with dedicated, demonstrated methods
Long single pulse (tens to 100+ ms)One flash, low heating, error strongly reduced100 ms may be insufficient for some large-format SHJ modules [2]; verify by forward/reverse test
Shaped voltage profiles in one pulseReproduced a multi-flash reference within ±0.5 % with a 10 ms pulse [6]Method must be validated per device type
Multi-flash (point or sector per flash)Each point at quasi-steady state; reference method in published comparisonsCycle time; flash-to-flash irradiance correction
Steady-state light (LED or lamp)Removes the capacitive artefact by constructionHeats the module; temperature control needed; slower
Residual capacitive error of step-like methods across five ISO/IEC 17025 laboratories and one industrial laboratory: (0.47 ± 0.19) % (k = 2) [7].

Where Pasan equipment fits

  • HighLIGHT 100ms Lab40 to 100 ms plateau, A+A+A+ to IEC 60904-9, with Pasan's DragonBack® method for highly capacitive modules (Pasan states a method uncertainty below 0.1 % with about ten times fewer flashes than standard multi-flash) and SmartSweep to choose the method per device.
  • HighLIGHT 100ms ProdThe same measurement core in production at about 15 s per module.
  • GridTOUCHTemporary contacting for busbarless and heterojunction cells.
  • SpotLIGHT cell testerSingle-cell I-V designed for capacitive cells.

A practical validation

  1. Measure the module with your production settings, sweeping forward and reverse.
  2. Repeat with a slower acquisition (longer pulse, multi-flash or steady state).
  3. If Pmax differs by more than your repeatability, lengthen the acquisition or change method until the results converge, then fix those settings for the product.

Independent evidence

  • Fast sweeps on back-contact and HIT modules under- or over-estimated Pmax by over 20 % depending on sweep time and direction; problems generally appear below about 200 ms acquisition time [1].
  • Even 100 ms simulators are no longer enough for accurate direct I-V of some recent n-type and SHJ modules [2].
  • A 10 ms single pulse with customised voltage profiles reproduced a multi-flash reference within ±0.5 % [6].
  • A single 50 ms pulse method for high-capacitance cells and modules has been reported [8].
  • Partial-curve scanning per flash has been proposed for in-line testing of high-efficiency cells [9].

Technical questions

Why do HJT modules show hysteresis on a flasher?
Because their junction and diffusion capacitances make the current lag a fast voltage sweep. Forward and reverse sweeps then err in opposite directions [1].
Is a 100 ms flash enough for HJT modules?
Often, not always. A 2021 review states that even 100 ms simulators are no longer enough for some current modules [2]. Verify on each product with a forward/reverse or two-speed test.
How do I know my HJT measurement is free of capacitive error?
Show equivalence to steady state, as IEC 60904-1:2020 requires: forward and reverse sweeps, or two acquisition times, must agree within your uncertainty target.
What residual error do accredited laboratories see?
About (0.47 ± 0.19) % (k = 2) with step-like methods in a 2019 round robin [7].
Is steady-state light always better for HJT?
It removes the capacitive artefact, but heats the module and takes longer. Long pulses with suitable methods can reach the same result in one flash when validated [6].

References

  1. European Commission JRC (N. Taylor, ed.), “Guidelines for PV power measurement in industry”, EUR 24359 EN, Publications Office of the European Union, 2010. Source
  2. M. Pravettoni et al., “The effect of capacitance on high-efficiency photovoltaic modules: a review of testing methods and related uncertainties”, Journal of Physics D: Applied Physics 54, 193001, 2021. doi:10.1088/1361-6463/abe574
  3. C. Monokroussos et al., “The effects of solar cell capacitance on calibration accuracy when using a flash simulator”, IEEE 4th World Conference on Photovoltaic Energy Conversion, 2006. doi:10.1109/WCPEC.2006.279953
  4. Q. Gao et al., “Effects of I-V measurement parameters on the hysteresis effect and optimization in high-capacitance PV module testing”, IEEE Journal of Photovoltaics 8, 710-718, 2018. doi:10.1109/JPHOTOV.2018.2810852
  5. M. Herman et al., “Optimal I-V curve scan time of solar cells and modules in light of irradiance level”, International Journal of Photoenergy 2012, 151452, 2012. doi:10.1155/2012/151452
  6. A. Virtuani, G. Rigamonti, G. Friesen, D. Chianese, P. Beljean, “Fast and accurate methods for the performance testing of highly-efficient c-Si photovoltaic modules using a 10 ms single-pulse solar simulator and customized voltage profiles”, Measurement Science and Technology 23, 115604, 2012. doi:10.1088/0957-0233/23/11/115604
  7. C. Monokroussos et al., “Electrical characterization intercomparison of high-efficiency c-Si modules within Asian and European laboratories”, Progress in Photovoltaics 27, 603-622, 2019. doi:10.1002/pip.3134
  8. H. Kojima et al., “Accurate and rapid measurement of high-capacitance PV cells and modules using a single short pulse light”, 40th IEEE Photovoltaic Specialists Conference, 2014. doi:10.1109/PVSC.2014.6925295
  9. A. Edler et al., “Understanding and overcoming the influence of capacitance effects on the measurement of high efficiency silicon solar cells”, Energy Procedia, 2012. doi:10.1016/j.egypro.2012.07.062
Related pages
TOPCon modulesBack-contact and IBCLong pulse vs steady stateGuide: xenon vs LED for HJT and TOPConArticle: dynamic I-V of capacitive modules

How this page is written: claims about physics and standards cite published sources, claims about Pasan equipment repeat only what the product pages state, and no other manufacturer is named. Editorial method

Rather we contact you?

Leave your email and we will get back to you. No form to fill in.

We only use your address to answer you. Privacy policy